A new electrochemical sensor based on Poly(Isoleucine) modified glassy carbon electrode decorated with platinum nanoparticles (Pt/Poly(Isoleucine)/GCE) was developed for sensitive individual and simultaneous determination of hydroquinone (HQ) and catechol (CC). Scanning electron microscopy (SEM), Electrochemical impedance spectroscopy (EIS), Cyclic voltammetry (CV) and Differential pulse voltammetry (DPV) were performed in order to characterize the Pt/Poly(Isoleucine)/GCE nanocomposite. For simultaneous determination of HQ and CC, Pt/Poly(Isoleucine)/GCE showed wide linear range between the 0.01-100.0 mu M. The detection limits were 0.006 mu M for HQ and 0.005 mu M for CC. The Pt/Poly(Isoleucine)/GC electrode exhibited good sensitivity and reliability in the simultaneous electroanalysis of two isomers in PBS of pH 7.5. The modified electrode was used to detect the isomers in naturel samples.